Skip to main content

Table 2 Sensitivity, negative predicted value (NPV) with confidence intervals (CI), McNemar and Kappa statistic of the different diagnostic tests used

From: Detection of Schistosoma mansoni DNA using polymerase chain reaction from serum and dried blood spot card samples of an adult population in North-western Tanzania

Test method

Sensitivity (95% CI)

NPV (95% CI)

P McNemar

Kappa (P value)

Microscopy (KK)

48.8% (37.9–59.9)

13.7% (5.7–26.3)

< 0.001

0.126 (P = 0.013)

POC-CCA

84.9% (75.5–91.7)

31.6% (12.6–56.6)

 = 0.002

0.395 (P < 0.001)

Serum-based real-time PCR

95.4% (88.5–98.7)

63.6% (30.8–89.1)

 = 0.125

0.755 (P < 0.001)

DBS real-time PCR

45.4% (34.6–56.5)

13% (5.4–24.9)

< 0.001

0.111 (P = 0.019)

  1. Serum-based real-time PCR and Kato-Katz results were used as reference (any positives by either of these two methods were deemed to be true positive)